Biological Safety Cabinets are a standard fixture in laboratories and controlled environments, designed to protect personnel, samples, and the surrounding environment. Traditionally, their role has been limited to handling biological agents under controlled airflow conditions.
But what if the same cabinet could do more?

With the addition of a mobile electric height adjustable base, Able Biological Safety Cabinets can be transformed into fully functional BioSafety Change Stations, offering a flexible solution for facilities looking to improve workflow, ergonomics, and space efficiency.
From Static Cabinet to Dynamic Workstation
A traditional Biological Safety Cabinet operates at a fixed height, which can limit usability across different staff members and tasks. By integrating a mobile electric base, the cabinet becomes adjustable, allowing users to position the work surface at a height that suits them. This approach not only improves flexibility and workflow, it also provides THE cost effective change station solution by combining two functions into a single piece of equipment.
This transforms the cabinet into a versatile change station suitable for:
- Animal handling and change procedures
- Cleanroom gowning support
- Sensitive workflow transitions
- Multi-user laboratory environments
Designed for Ergonomics and Accessibility
The actuator system offers a minimum compression height of 584 mm and a stroke length of 650 mm, providing a wide adjustment range that is rarely matched in the industry.
- Minimum work surface height: approximately 884 mm
- Maximum work surface height: up to 1534 mm
- Overall unit height range: approximately 2384 mm to 3034 mm
This range allows both shorter and taller technicians to work comfortably, reducing strain and supporting better posture during extended use.
Maintaining Certified Protection
Even with its added flexibility, the core function of the cabinet remains unchanged. Able Biological Safety Cabinets are:
- EN 12469 compliant for use across Europe
- Will be designed to meet AS2252.2-2025 requirements for Australia and New Zealand
This ensures that whether used as a traditional cabinet or a change station, the unit continues to deliver the required level of protection.
Two Cabinet Options to Suit Different Environments
Metal Class II Type A2 Biological Safety Cabinet
Constructed with an epoxy coated metal exterior and SS304 internal surfaces, this model offers durability and reliability for general laboratory use.
Key features include:
- Dual HEPA H14 filters with 99.9995% efficiency at 0.3 µm
- 70% downflow and 30% exhaust airflow pattern
- 10.1 inch touchscreen control system
- UV sterilisation system with safety interlock
- Real time monitoring and alarm systems
- ISO 5 / Class 100 cleanliness level
Polypropylene Class II Type A2 Biological Safety Cabinet
Designed for environments requiring higher chemical resistance, this model features a fully welded polypropylene structure with excellent durability against acids and corrosive agents.
Key features include:
- Dual ULPA U15 filters with 99.9995% efficiency at 0.1 µm
- Integrated Particle Monitoring System with contamination alarm
- Electric lift front window and side visibility panels
- Stainless steel work surface
- Advanced monitoring, alarms, and environmental display
Why Convert to a Change Station?
Combining a Biological Safety Cabinet with a height adjustable base offers several advantages:
- Improved ergonomics for all users
- Greater flexibility across different applications
- Reduced need for separate equipment
- Better use of available space
- Consistent contamination control during change procedures
A Smarter Use of Equipment
The ability to convert a Biological Safety Cabinet into a BioSafety Change Station reflects a broader shift towards adaptable laboratory infrastructure.
Instead of dedicating equipment to a single purpose, facilities can maximise functionality while maintaining compliance and performance standards.
In the end, it raises a simple question, is it just a Biological Safety Cabinet, or is it something more?
Contact our consultants for more information.